gratingcalibration

Tool for calibrating a SAXS beamline using diffraction grating. Currently under construction.

What Where
Source https://github.com/DiamondLightSource/i22-diffraction-grating-calibration
Docker docker run ghcr.io/diamondlightsource/i22-diffraction-grating-calibration:latest
Releases https://github.com/DiamondLightSource/i22-diffraction-grating-calibration/releases

Installation

On the DLS module system, you can install this tool as:

module load uv
uv venv 
uv pip install git+https://github.com/DiamondLightSource/i22-diffraction-grating-calibration.git
source .venv/bin/activate

Usage

$ i22-diffraction-grating-calibration --file /path/to/input.nxs --output /path/to/output/folder --save-plots

The calibration NeXuS file (SAXS_calibration.nxs) will be stored in the output_folder, along with plots made to show how the calibration has been performed.

The final figure generated, detector_calibration.png, shows how peaks have been found, fit, and indexed in the azimuthally integrated scattering pattern to calculate the detector distance from the sample. The initial peak identification is performed using scipy.signal.find_peaks, using a peak prominance of 0.1. This has performed well in testing, but may not be universally optimal. Should this need fine tuning, the following command could be used instead:

$ i22-diffraction-grating-calibration --file /path/to/input.nxs --output /path/to/output/folder --save-plots --peak-prominance X

where X is some float value . Set this lower (than 0.1 to find more peaks, higher to find fewer.

What's happening?

This calibration program works by:

  1. Identifying the beamstop & fitting and approximate center
    • Shown in the beamstop_fit.png figure
  2. Optimising the beam centre through x/y detector profile matching
    • See beam_profiles_xy.png and beam_center_location.png to see how well the profiles have been matched, and where the beam center has been located
  3. Calibrate the detector distance based on grating fringe spacing
    • The peaks found, and how they have been indexed, is shown in detector_calibration.png.

Notes

  • Assumes a Pilatus 2M detector, pixel size 172e-6 m.
  • Units are all SI, not more native/sensible SAXS based units (e.g. beam center position should really be in mm not m, wavelength in angstrom not m, etc.)